AccuSim Monitor Simulator,Handheld AccuSim Total Patient Simulator

Mode:AccuSim

Type:QA-Instruments

Contact:   ken@hkmedqc.com

The AccuSim Total Patient Simulator offers a synchronized simulation system that includes invasive blood pressure, breathing, and temperature functions. It combines our unparalleled NIBP and SpO2 technologies within a powerful and convenient handheld device.

The AccuSim monitor simulator and AccuSim monitor detector accurately simulate, test, and calibrate all patient monitors and have been recognized by Masimo, Welch-Allyn, and other major manufacturers.
Our measures were developed in collaboration with the world's leading patient monitor manufacturers. The AccuSim is the only simulator with such a wide range of monitoring functions.
A fully synchronized patient monitor simulation system
Built-in pressure gauge and high-precision pressure sensor
Accurate NIBP simulation using CalTables
Built-in adult non-invasive blood pressure cuff
Precise second-generation SpO2 simulation with R curve
Precise testing of Masimo's Rainbow technology and other new pulse oximetry measurement techniques
12-lead electrocardiogram simulation
Invasive blood pressure simulation
Respiratory simulation
Temperature simulation
Arrhythmias and pacemakers
USB connection
Internet-updatable firmware
Pre-set SpO2 R curve for easier monitoring
 
Measurement range: 10 - 295 mmHg;
Measurement accuracy: ±0.5 mmHg
Application: A detection instrument integrating multi-parameter simulation functions such as non-invasive blood pressure, SpO2, electrocardiogram, etc.

AccuSim Monitor Simulator,Handheld AccuSim Total Patient Simulator

Product features of AccuSim multi-parameter monitor and comprehensive tester:

Ø Integrates multiple parameter simulation functions such as non-invasive blood pressure, blood oxygen saturation, and electrocardiogram;
Ø Equipped with both optical and electrical simulation functions for blood oxygen saturation;
Ø Offers optional non-invasive blood pressure algorithms simulation (CalTablesTM) from over ten major manufacturers including GE, Philips, and Spacelabs;
Ø Certified by manufacturers with high-precision blood oxygen R curves (including Philips, Omeda, Masimo, Nellcor, etc.);
Ø Blood pressure simulation accuracy: ±0.5 mmHg; Technical Specifications
Ø Non-invasive Blood Pressure
Pre-set 7 sets of adult and 6 sets of neonatal blood pressure simulation values;
Blood pressure simulation range: 10 - 295 mmHg;
Heart rate range: 15 - 330 BPM;
Pulse amplitude: 0 - 150%
Static pressure range: 0.0 - 400.0 mmHg;
Accuracy: ±0.5 mmHg;
Built-in air pump, supporting pressure leakage, overpressure protection tests and digital pressure gauge functions
 
Ø Blood Oxygen Saturation
Pre-set 14 sets of blood oxygen R curves;
Range: 0 - 100 SPO2%;
Accuracy: ±0.5 SPO2% (70 - 100 SPO2%), ±1 SPO2% (50 - 69 SPO2%);
Perfusion amplitude: 0.10 - 20.0%
Can add environmental light, power frequency/frequency doubling, movement, arrhythmia, etc. as interference;
 
Ø Electrocardiogram
Heart rate: 30 - 330 BPM;
Amplitude: 0.15 - 5.00 mV;
Supports parameter adjustments for pacing, S-T segment, T wave, R-R interval, etc.;
Arrhythmias: tachycardia, bradycardia, ventricular tachycardia, atrial fibrillation, ventricular fibrillation, asystole, etc.;
 
Ø Respiration
Respiratory rate: 4 - 120 BPPM;
Baseline impedance: 250, 500, 750, 1000 Ω;
 
Ø Temperature
Range: 0.0 - 50.0℃;
 
Ø Invasive Blood Pressure
Static pressure: 0 - 300 mmHg (step 1 mmHg);
 

Overall dimensions

4.25 inches wide x 2 inches high x 11.5 inches long
Including the built-in adult cuff support

AccuSim Monitor Simulator,Handheld AccuSim Total Patient Simulator


 SAG: AccuSim,AccuSim monitor tester,AccuSim total patient simulator,Clinical Dynamics AccuSim,patient monitor simulator,non-invasive bp tester,invasive blood pressure simulator,SpO2 simulator,respiration temperature simulator,handheld patient simulator