ESU-2400H is a high-end upgraded electro-surgical analyzer. Its hardware platform is shared with ESU-2400. The old ESU-2400 equipment can be upgraded to the H version through official authorization without the need to replace the entire hardware. It is equipped with the patented DFA2 high-speed waveform digitization technology, which can accurately capture the multi-amplitude complex pulse radio frequency output of the new electrosurgical device. It can measure the spray, mixed mode, and multi-pulse complex surgical waveforms with stable and reliable results. It features a 5.7-inch color touchscreen, with an internal 0-6400Ω non-sensory precise load, and a 1Ω step-adjustable feature. Complete test items include output power, radio frequency voltage and current, peak factor, high-frequency leakage current, REM/ARM circuit electrode monitoring, duty cycle, pulse timing, and automatic power-load curve scanning. It is widely used in hospital equipment departments, third-party metrology calibration laboratories, and the original equipment manufacturer's research and development and verification of electrosurgical devices.

Core technical parameters
▪ Display: 5.7-inch QVGA color touch screen
▪ Built-in non-sensing load: 0 - 6400Ω, 1Ω step, ±1% accuracy
▪ Measurement frequency: 10 kHz - 10 MHz
▪ Power range: 0 - 999.9 W, resolution 0.1 W
▪ Maximum input voltage: 10 kV peak
▪ Current range: 2 mA - 7000 mA RMS
▪ Peak factor: 1.4 - 500
▪ Core features: Patent DFA2 high-speed digitization, supports up to 3 levels of amplitude multi-pulse RF waveform measurement
▪ Measurement items: Output power, RF voltage current, peak factor, duty cycle, pulse timing, high-frequency leakage current, REM/ARM negative plate monitoring test
▪ Interfaces: 3×USB, network port, serial port, BNC waveform output, PS/2, foot switch interface
▪ Data: Test report export, automatic sequence test, waveform screenshot saving
▪ Power supply: Wide voltage AC 100 - 240V 50 - 60Hz
▪ Dimensions: 198.1 × 381 × 571.5 mm; Weight 14 kg
| Comparison Item |
ESU-2400H |
ESU-2400 |
| Core Technology |
Patented DFA2 high-speed digitization engine |
Standard DFA digitization |
| Multi-Pulse Measurement |
Supports up to 3-amplitude complex pulsed RF waveforms, for latest modern ESU generators |
Only single-amplitude pulse / continuous wave; cannot resolve multi-level pulse output |
| Measurement Parameters |
Up to 28 selectable measurement parameters |
Up to 15 selectable measurement parameters |
| ForceTriad / Ligasure OEM auto-sequence |
Full native support |
Basic support, limited multi-pulse test capability |
| Upgrade Path |
Can be upgraded from existing ESU-2400 hardware via license upgrade |
Can upgrade to ESU-2400H with official license key |
| Screen |
5.7-inch color touch screen |
5.7-inch color touch screen |
| Internal Load Bank |
0-6400Ω @1Ω step |
0-6400Ω @1Ω step |
| Typical Target Customer |
Metrology calibration laboratory, testing new multi-pulse electrosurgical knife, equipment manufacturer research and development |
Hospital Equipment Department, Routine Preventive Maintenance of Traditional Continuous Wave Electrosurgical Knife |
Product Highlights
✔ Patent DFA2 high-speed sampling, stable and reliable measurement readings for spray, mixing mode, and complex multi-pulse radio frequency waveforms
✔ Three-level multi-pulse waveform analysis, fully compatible with Valleylab ForceTriad, Ligasure, and other new-generation intelligent electrosurgical equipment
✔ Complete set of built-in non-sensory load, 0-6400Ω with 1Ω fine adjustment, automatic power-load curve scanning
✔ Complete set of tests including power, peak factor, leakage current, REM negative plate monitoring, and pulse timing in one unit
✔ Network port + USB data export, built-in automatic test sequence, improves quality control efficiency, and meets compliance report archiving requirements
✔ Consistent with ESU-2400 hardware, old equipment can be directly purchased and authorized upgraded without the need to replace the entire machine
✔ BNC waveform monitoring output, external oscilloscope for intuitive observation of radio frequency waveforms