HF voltage, RMS
High-frequency output power: During the power measurement process, the software first sets the specified load resistance to 10 ohms or 25 ohms, ranging from 25 ohms to 6375 ohms in steps of 25 ohms. The HF's output power can be sent to HF-400 for measurement. An automatic range switcher is responsible for the optimal control of the RMS converter. This RMS converter, based on the thermal conversion principle and in coordination with the driver module, is specifically designed to reach 10 MHz.
High-frequency leakage current:
The high-frequency leakage current is measured through a 200 ohm load resistor. For this test, the load resistor is adjustable.
Main features
Full-function ESU tester compliant with IEC 60601-2-2 safety standard
Test items: High-frequency power, high-frequency RMS voltage/current, high-frequency leakage current, peak factor, neutral electrode test
Adjustable electronic load resistance: 10 ohms, 25–6375 ohms, with a step size of 25 ohms; 6 selectable preset resistance values
Measurement range: High-frequency power 0–500 watts; High-frequency leakage current 0–250 milliamps; High-frequency current 0–5000 milliamps
Accuracy: Power ±1 watt or ±2.5% of reading; Leakage current ±2 milliamps or ±5% of reading
Dual operation modes: Local menu-driven independent operation or automatic control via PC
Interfaces: USB and RS-232 for PC data recording, report generation, and automated test sequence
4×20 LCD display; supports multilingual menu
Portable rugged desktop design, equipped with an integrated handle for on-site service