NIM-ECG-1 Digital ECG EEG Calibrator for Electrocardiograph,Brain Electrograph Details:
Built-in point procedures and points, real-time display of ECG waveform
The digital electrocardiogram machine calibration instrument is a precise instrument capable of simulating various signal waveforms. It can be used for the calibration and verification of electrocardiogram machines, electroencephalogram machines, digital electrocardiogram machines, electroencephalogram machines, digital electroencephalogram machines, cardiac defibrillators (electrical part, dynamic electrocardiogram machines, electrocardiogram monitors, exercise treadmill machines, etc.).
Built-in regulations: JJG 543-2026 "Electrocardiograph", JJG 760-2003 "Cardiac Monitoring Instrument", JG 1042-2008 "Dynamic (Portable) Electrocardiograph", JJG 954-2019 "Digital Electroencephalograph", JJG 1043-2008 "Electroencephalograph", JJF 1149-2014 "Calibration Specification for Cardiac Defibrillator" (Electrocardiogram Section), JJG 749-2007 "Cardiac and Brain Electrograph Instrument", JJG 1016-2006 "Calibration Instrument for Electrocardiograph and Brain Electrograph", JJF 1722-2018 "Calibration Specification for Exercise Treadmill"; Other applicable regulations include: JJG 1041-2008 "Digital Electrocardiograph", JJG 1163-2018 "Multi-parameter Monitor" (Electrocardiogram Section)

Technical Features
The modular interface is designed according to standard specifications, with clear and understandable processes.
It features a 10.1-inch high-definition touch screen that can display real-time output waveform signals. It also includes a built-in procedure guide.
The operation is simple and convenient.
It integrates the required AC signal generator component for dynamic electrocardiogram machine calibration.
It is equipped with a rechargeable battery that can operate continuously for over 10 hours after a single charge.
It comes with an independent common-mode suppression ratio calibration device, which can display the adjusted output voltage through the built-in display screen.
Technical Specification
1. Square wave signal: Voltage peak-to-peak (0.001mV - 14V), maximum allowable error in (0.5mV - 14V): ±1%; Frequency (0.001Hz - 1000Hz), maximum allowable error in (0.1Hz - 1000Hz): ±1%; Duty cycle (0.1 - 0.9); Input impedance: < 600Ω
2. Sine wave: Voltage peak-to-peak (0.001mV - 28V), maximum allowable error in (0.5mV - 14V): ±1%; Frequency (0.001Hz - 1000Hz), maximum allowable error in (0.1Hz - 1000Hz): ±1%; Input impedance: < 600Ω; Distortion: < 5%
3. Triangle wave: Voltage peak-to-peak (0.001mV - 14V) and (-0.001mV - -14V), maximum allowable error in (0.5mV - 14V) and (-0.5mV - -14V): ±1%; Frequency (0.001Hz - 1000Hz), maximum allowable error in (0.1Hz - 1000Hz): ±1%
4. Differential signal: Time constant (0.01ms - 1000ms), period (1ms - 100000ms), maximum amplitude (0.1mV - 12V), pulse time (0ms - 1000ms)
5. Polarization voltage: ±150mV, ±300mV, ±600mV, ±950mV, ±1000mV, ±2000mV, ±3000mV, ±4000mV; Maximum allowable error at ±300mV point: ±0.5%
6. Analog skin-electrode impedance Z1: 51kΩ in parallel with 47nF capacitor, resistance maximum allowable error: ±1%, capacitor maximum allowable error: ±5%
7. Input impedance calibration sampling impedance Z2: 620kΩ resistor in parallel with 4700pF capacitor, resistance maximum allowable error: ±1%, capacitor maximum allowable error: ±5%
8. Input loop current sampling resistor: 10kΩ, maximum allowable error: ±5%
9. R01 is 100kΩ, with a maximum allowable error of ±0.1%; R02 is 100Ω, with a maximum allowable error of ±0.1%; RA1 is all 100kΩ, with a maximum allowable error of ±1%; R03 is 1kΩ, with a maximum allowable error of ±1%; R04 is 111Ω, with a maximum allowable error of ±1%.
10. CA1 is 0.5uF, with a maximum allowable error of ±5%.
11. Attenuator: 1000:1, with a maximum allowable error of ±0.1%.
12. ECG simulation signal: (1~380) times/min, with a maximum allowable error of ±0.1%; amplitude: the peak-to-valley range of the output signal amplitude is (0.002mV~28V); when the peak-to-valley amplitude is 1mV, the maximum allowable error is ±1.0%; when the peak-to-valley amplitude is 0.3mV~10.0mV, the maximum allowable error is ±1.5%; when the peak-to-valley amplitude is 0.1mV~0.2mV, the maximum allowable error is ±3.0%; when the peak-to-valley amplitude is 0.03mV~0.05, the maximum allowable error is ±10%.
13. Heart rate signal: (1~350) times/min, with a maximum allowable error of ±1.0%; amplitude (0.1mV~14V), with a maximum allowable error not exceeding ±2%.
14. Pacing pulse waveform (sinus wave superimposed with pulse wave signal): sinus wave 2mV, 10Hz, pulse wave 2mV, period 0.6s; sinus wave 2mV, 10Hz, pulse wave 200mV, period 0.6s; sinus wave 2mV, 10Hz, pulse wave 20mV, period 0.6s.
15. Common mode rejection ratio verification device: monitoring voltage meter measurement range: 9V~11V (effective value), maximum allowable error: ±10%; input impedance: ≥300MΩ; frequency: 50Hz; equivalent output voltage: 10V (effective value); equivalent output capacitance: 200pF.
16. Other signals: can simulate "pulse wave, maternal-fetal electrocardiogram mixed signal, synthetic electrocardiogram wave, enhanced synthetic electrocardiogram wave, enhanced synthetic electrocardiogram wave, computer-generated random noise, DC offset, trapezoidal signal, DC Gaussian noise, double rhythm_3a, slow-changing double rhythm_3b, fast-changing double rhythm_3c, bidirectional contraction_3d, ventricular tachycardia_1mV_206bpm_4a, ventricular tachycardia_2mV_195bpm_4d".
17. Adopting a 10.1-inch high-definition touch screen, with built-in procedure guide and data processing, built-in rechargeable battery, equipped with an independent common mode rejection ratio verification device, and can display the adjusted output voltage through the built-in display screen.
Product Configuration
General: Digital Electroencephalogram Analyzer, Common Mode Suppression Ratio Testing Device, Capacitor Adjustment Insulation Screwdriver, Power Adapter, USB Cable, Protective Case.
Optional Accessories: Steel Ruler, Divider, Magnifying Glass, Stopwatch
SAG: ECG calibrator,EEG calibrator,ECG EEG tester,electrocardiograph calibrator,electroencephalograph calibrator,ECG simulator,EEG signal generator,medical ECG calibration device,electrocardiogram tester,electroencephalogram calibrato