| Item | Specifications |
| Square Wave | Amplitude: (0.005 ~ 5.0) mV; Maximum permissible error: ±2% within (0.5 ~ 2.0) mV.Frequency: (0.001 ~ 2000) Hz; Maximum permissible error: ±1% within (0.2 ~ 2.5) Hz.Duty cycle: 0.1 ~ 0.9 |
| Sine Wave | Amplitude: (0.005 ~ 5.0) mV; Maximum permissible error: ±2% within (0.5 ~ 2.0) mV.Frequency: (0.001 ~ 2000) Hz; Maximum permissible error: ±1% within (0.1 ~ 100) Hz |
| Standard Heart-Rate Signal | Amplitude: (0.1 ~ 5.0) mV; Maximum permissible error: ±2% within (0.5 ~ 2.0) mV.Heart rate: (10 ~ 360) bpm; Maximum permissible error: ±(1 % of reading + 1) bpm for (30 ~ 300) bpm |
| Pulse Wave | Amplitude: (0.005 ~ 5) mV; Period: (0.001 ~ 100) s;Pulse width: (0.1 ~ 50000) ms; Delay: (0 ~ 50000) ms |
| Triangle Wave | Amplitude: (0.005 ~ 5) mV; Period: (0.001 ~ 1000) s;Rise time: (0.1 ~ 500000) ms; Fall time: (0.1 ~ 500000) ms |
| Maternal-Fetal ECG Mixed Signal | Maternal amplitude: (1 ~ 2) mV; Maternal heart rate: (10 ~ 360) bpm.Fetal amplitude: (0 ~ 1) mV; Fetal heart rate: (10 ~ 360) bpm |
| Synthesized ECG Waveform | Heart rate: (10 ~ 360) bpm; R-wave amplitude: (0 ~ 2) mV; S-wave amplitude: (0 ~ −2) mV;ST-segment baseline amplitude: (−2 ~ 2) mV; T-wave amplitude: (0 ~ 2) mV; P-wave amplitude: (0 ~ 2) mV;QRS-wave amplitude: (0 ~ 2) mV; QRS-wave interval: (0 ~ 750) ms; ST-wave interval: (0 ~ 750) ms;T-wave interval: (0 ~ 750) ms; P-wave interval: (0 ~ 750) ms; PQ-wave interval: (0 ~ 750) ms |
| Computer-Generated Random Noise | (0 ~ 2) mV |
| ECG Simulation Signal | Heart rate, P1 amplitude, P2 amplitude, Q amplitude, R1 amplitude, R2 amplitude, R3 amplitude, R4 amplitude, ST amplitude, T amplitude, P1 time, P2 time, PQ time, R1 time, R2 time, R3 time, R4 time, S time, ST time, T time, QT time. All parameters are independently adjustable. |
| Differential Signal | Period: (0 ~ 10000) ms; Amplitude: (0 ~ 2) mV; Time constant: (0.01 ~ 1000) ms;Negative-pulse duration: (0 ~ 10000) ms |
| Other Signals | Simulates DC offset, trapezoidal signal, DC Gaussian noise, bigeminy_3a, slow-varying bigeminy_3b, fast-varying bigeminy_3c, bidirectional contraction_3d, ventricular tachycardia_1mV_206bpm_4a, ventricular tachycardia_2mV_195bpm_4b and other waveforms. |
| Item | Specifications |
| Pressure Units | mmHg, kPa, cmH₂O, inH₂O, psi |
| Pressure Measurement | Measuring range: (0 ~ 400) mmHg; Resolution: 0.1 mmHg; Maximum permissible error: ≤ ±0.8 mmHg |
| Pressure Source | Output range: (0 ~ 400) mmHg; Resolution: 1 mmHg.Deviation between target and actual pressure: ±10 mmHg for 100-400 mmHg with minimum volume of 300 cc. |
| Dynamic Blood-Pressure Simulation (Oscillometric Method) | Systolic pressure: (25 ~ 255) mmHg; Diastolic pressure: (15 ~ 245) mmHg;Blood-pressure repeatability: ≤ 2.25 mmHg; Heart-rate range: (30 ~ 200) bpm;Maximum heart-rate error: ±(1 % reading + 1) bpm;Cuff types: external cuff, internal cuff |
| Custom Blood-Pressure Simulation | Presets: 7 standard blood-pressure profiles, 8 patient states, 4 arrhythmia modes, 4 respiration modes, 4 neonatal modes, 3 wrist-cuff modes |
| Air-Tightness Verification | Pre-configured verification workflow per regulatory requirements. System pressure points are read at 1-minute and 5-minute intervals, and pressure-leakage rate is calculated automatically. |
| Item | Specifications |
| Pulse Oxygen Saturation | Simulation range: (30 % ~ 100 %); Resolution: 1 %; Repeatability: ≤ 1 %;Maximum permissible error: ±2 % for 75 %-100 %; ±3 % for 35 %-74 % |
| Pulse Rate | Simulation range: (25 ~ 300) bpm; Resolution: 1 bpm;Maximum permissible error: ±1 % of reading or ±1 bpm, whichever is larger |
| Perfusion Index | Range: (0.000 % ~ 20.000 %);Resolution: 1 % within (1.000 %-20.000 %); 0.1 % within (0.100 %-0.900 %) |
| 10 built-in R-curves: BCI, Criticare, Datascope, Datex, PMS M1190, Masimo, Nellcor, Nihon-Kohden, Ohmeda&Nova, Respironics | |
| Item | Specifications |
| Respiration rate | (0 ~ 150) breaths/min; Accuracy: ±0.5 % |
| Gas Interfaces | -CO₂calibration‑gas input: quick‑connect fitting for 6 mm‑outer‑diameter PU tubing - Modulated-gas output: barb fitting, outer diameter 15 mm, length 20 mm |
| Matching Accessories - CO₂ calibration gas: 5 % volume fraction (nitrogen balance gas); Maximum permissible error: ±2 % - Two‑stage pressure‑reducing valve (with gas flowmeter): High‑pressure gauge range 25 MPa; Low‑pressure gauge range 0.6 MPa; CO₂ outlet pressure: (0.15 ~ 0.2) MPa - Precision digital barometer: 3.5‑inch display with numerical and curve readout; Pressure measuring range 900‑1100 hPa; Accuracy class 0.1; Units: kPa, hPa, Pa, mmHg configurable. |
|