NIMM-PS-2 Multi-Parameter Physiological Patient Simulator

Mode:NIMM-PS-2

Type:QA-Instruments

Contact:   ken@hkmedqc.com

NIMM-PS-2 Multi-Parameter Physiological Patient Simulator

Built-in point procedures and positions are more in line with the operation habits of domestic first-line users.
It is equipped with a detachable internal lithium battery and adopts a modular design. The main unit features a 7-inch touch-screen color LCD display, allowing multiple physiological parameters to be displayed simultaneously. It also includes internal calibration procedures for multi-parameter monitors, electronic blood pressure meters, mercury blood pressure meters, and blood pressure charts.
Measurements can be entered on the screen, and the errors or repeatability can be automatically calculated. Electronic versions of the original records and certificates can be issued based on the original data recorded by the inspector, and they can also be exported.
It integrates 8 physiological parameter simulation functions such as electrocardiogram (ECG), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), end-tidal carbon dioxide (EtCO2), cardiac output (CO), body temperature (TEMP), respiration (chest impedance method simulation), and pulse oxygen saturation (SPO2).

NIMM-PS-2 Multi-Parameter Physiological Patient Simulator

Technical Specification

-ECG (Electrocardiogram) Section

Item Specifications
Square Wave Amplitude: (0.005 ~ 5.0) mV; Maximum permissible error: ±2% within (0.5 ~ 2.0) mV.Frequency: (0.001 ~ 2000) Hz; Maximum permissible error: ±1% within (0.2 ~ 2.5) Hz.Duty cycle: 0.1 ~ 0.9
Sine Wave Amplitude: (0.005 ~ 5.0) mV; Maximum permissible error: ±2% within (0.5 ~ 2.0) mV.Frequency: (0.001 ~ 2000) Hz; Maximum permissible error: ±1% within (0.1 ~ 100) Hz
Standard Heart-Rate Signal Amplitude: (0.1 ~ 5.0) mV; Maximum permissible error: ±2% within (0.5 ~ 2.0) mV.Heart rate: (10 ~ 360) bpm; Maximum permissible error: ±(1 % of reading + 1) bpm for (30 ~ 300) bpm
Pulse Wave Amplitude: (0.005 ~ 5) mV; Period: (0.001 ~ 100) s;Pulse width: (0.1 ~ 50000) ms; Delay: (0 ~ 50000) ms
Triangle Wave Amplitude: (0.005 ~ 5) mV; Period: (0.001 ~ 1000) s;Rise time: (0.1 ~ 500000) ms; Fall time: (0.1 ~ 500000) ms
Maternal-Fetal ECG Mixed Signal Maternal amplitude: (1 ~ 2) mV; Maternal heart rate: (10 ~ 360) bpm.Fetal amplitude: (0 ~ 1) mV; Fetal heart rate: (10 ~ 360) bpm
Synthesized ECG Waveform Heart rate: (10 ~ 360) bpm; R-wave amplitude: (0 ~ 2) mV; S-wave amplitude: (0 ~ −2) mV;ST-segment baseline amplitude: (−2 ~ 2) mV; T-wave amplitude: (0 ~ 2) mV; P-wave amplitude: (0 ~ 2) mV;QRS-wave amplitude: (0 ~ 2) mV; QRS-wave interval: (0 ~ 750) ms; ST-wave interval: (0 ~ 750) ms;T-wave interval: (0 ~ 750) ms; P-wave interval: (0 ~ 750) ms; PQ-wave interval: (0 ~ 750) ms
Computer-Generated Random Noise (0 ~ 2) mV
ECG Simulation Signal Heart rate, P1 amplitude, P2 amplitude, Q amplitude, R1 amplitude, R2 amplitude, R3 amplitude, R4 amplitude, ST amplitude, T amplitude, P1 time, P2 time, PQ time, R1 time, R2 time, R3 time, R4 time, S time, ST time, T time, QT time. All parameters are independently adjustable.
Differential Signal Period: (0 ~ 10000) ms; Amplitude: (0 ~ 2) mV; Time constant: (0.01 ~ 1000) ms;Negative-pulse duration: (0 ~ 10000) ms
Other Signals Simulates DC offset, trapezoidal signal, DC Gaussian noise, bigeminy_3a, slow-varying bigeminy_3b, fast-varying bigeminy_3c, bidirectional contraction_3d, ventricular tachycardia_1mV_206bpm_4a, ventricular tachycardia_2mV_195bpm_4b and other waveforms.

NIBP (Non-Invasive Blood Pressure) Section


Item Specifications
Pressure Units mmHg, kPa, cmH₂O, inH₂O, psi
Pressure Measurement Measuring range: (0 ~ 400) mmHg; Resolution: 0.1 mmHg; Maximum permissible error: ≤ ±0.8 mmHg
Pressure Source Output range: (0 ~ 400) mmHg; Resolution: 1 mmHg.Deviation between target and actual pressure: ±10 mmHg for 100-400 mmHg with minimum volume of 300 cc.
Dynamic Blood-Pressure Simulation (Oscillometric Method) Systolic pressure: (25 ~ 255) mmHg; Diastolic pressure: (15 ~ 245) mmHg;Blood-pressure repeatability: ≤ 2.25 mmHg; Heart-rate range: (30 ~ 200) bpm;Maximum heart-rate error: ±(1 % reading + 1) bpm;Cuff types: external cuff, internal cuff
Custom Blood-Pressure Simulation Presets: 7 standard blood-pressure profiles, 8 patient states, 4 arrhythmia modes, 4 respiration modes, 4 neonatal modes, 3 wrist-cuff modes
Air-Tightness Verification Pre-configured verification workflow per regulatory requirements. System pressure points are read at 1-minute and 5-minute intervals, and pressure-leakage rate is calculated automatically.

SpO₂(Pulse Oximetry) Section

Item Specifications
Pulse Oxygen Saturation Simulation range: (30 % ~ 100 %); Resolution: 1 %; Repeatability: ≤ 1 %;Maximum permissible error: ±2 % for 75 %-100 %; ±3 % for 35 %-74 %
Pulse Rate Simulation range: (25 ~ 300) bpm; Resolution: 1 bpm;Maximum permissible error: ±1 % of reading or ±1 bpm, whichever is larger
Perfusion Index Range: (0.000 % ~ 20.000 %);Resolution: 1 % within (1.000 %-20.000 %); 0.1 % within (0.100 %-0.900 %)
10 built-in R-curves: BCI, Criticare, Datascope, Datex, PMS M1190, Masimo, Nellcor, Nihon-Kohden, Ohmeda&Nova, Respironics

EtCO₂ (End-Tidal Carbon Dioxide) Section [Optional]

Item Specifications
Respiration rate  (0 ~ 150) breaths/min; Accuracy: ±0.5 %
 
Gas Interfaces -CO₂calibration‑gas input: quick‑connect fitting for 6 mm‑outer‑diameter PU tubing
- Modulated-gas output: barb fitting, outer diameter 15 mm, length 20 mm
 
Matching Accessories
 
- CO₂ calibration gas: 5 % volume fraction (nitrogen balance gas); Maximum permissible error: ±2 %
- Two‑stage pressure‑reducing valve (with gas flowmeter): High‑pressure gauge range 25 MPa; Low‑pressure gauge range 0.6 MPa; CO₂ outlet pressure: (0.15 ~ 0.2) MPa
- Precision digital barometer: 3.5‑inch display with numerical and curve readout; Pressure measuring range 900‑1100 hPa; Accuracy class 0.1; Units: kPa, hPa, Pa, mmHg configurable.
 

Product Configuration

 
Main unit, power adapter, oximetry simulation finger, temperature‑measurement lead set, respiration‑rhythm generator and accessories (optional).







 SAG: NIMM-PS-2,NIMM-PS-2 multi-parameter patient simulator,vital sign simulator,physiological simulator,patient monitor tester,multi-parameter physiological simulator,ECG simulator,NIBP simulator