HANS RUDOLPH 1120 Flow Simulator and 1120 Spirometry Check Waveform Generator are servo motor-driven piston pumps, which can be used for testing pulmonary function products and other respiratory equipment. They are designed for product development and manufacturing test applications. The flow/volume simulator has three working modes. The exhalation waveform mode is used to test spirometers using ATS waveforms, peak flow waveforms, or custom waveforms. The steady-state flow mode can generate steady-state flow within a wide range of flow rates. This mode can be used for testing and calibrating flow meters. The breathing waveform mode can generate continuous inhalation/exhalation flow waveforms that simulate breathing. All modes provide graphical representations of flow and pressure signals. Data collected during testing can be saved to a file for additional analysis.
The built-in pressure sensor measures air pressure and cylinder pressure during testing. Pressure data is used to correct the calculated flow for gas compression. Another differential pressure sensor is used to connect to the pressure faucet on or near the tested equipment. This pressure reading can be used to determine the flow resistance of the tested equipment. An optional Hans Rudolph Company gas cutter can be connected to the outlet of the flow/volume simulator. It provides a pressure sensor to measure the pressure difference on the pneumatic instrument, so that the measured flow signal can be plotted and recorded.
Two graphs are used to display the collected flow and pressure data. The flow graph can show the flow calculated based on the piston position, the flow measured by the air pump, and the target flow calculated from the required waveform. The pressure graph can show cylinder pressure and auxiliary pressure signals. Both graphs can be dynamically scaled and translated to allow users to zoom in and view specific parts of the waveforms or view the complete waveforms. Graphs can be printed.
Spirometry testing
In the exhalation waveform mode, the flow/volume simulator will generate all ATS waveforms for testing pulmonary function products. Custom waveforms can be created using the provided utility. Waveforms can be scaled up or down proportionally to adjust waveforms for different flow and capacity. Different piston starting positions allow you to change the internal volume.
Peak flow meter testing
Waveforms defined by EN13826 can be generated for testing peak flow meters. This simulator has been tested by the German Federal Technical Commission (PTB).
Steady flow
A steady flow state can be used to verify the calibration of the flowmeter and also provide a constant flow for the calibration of the flowmeter. The volume being transported can be adjusted to 8.5 liters. The required flow rate can be set from 0.1 to 16 liters per second. You can choose the direction of the airflow, either inhalation or exhalation. The pressure and flow data collected during the test are displayed on the graph after the test is completed.
Respiratory waveform
A respiratory waveform can be created using the provided utility, and then used to create a simulated breathing flow pattern. The amplitude and respiratory rate are adjustable, so you can simulate different breathing with the same waveform file. The respiratory waveform can be run once or repeated continuously.
Features of the 1120 Spirometry Check Waveform Generator:
• Produces all ATS spirometry measurement waveforms and peak flowmeter waveforms
• Lung volume measurement, steady-state flow, and continuous breathing waveform modes
• Tool for generating user-defined waveforms
• Graphical display of flow and pressure signals
• Saving of data files for further analysis
• Built-in pressure sensor for measuring pressure drop of the tested equipment, ±10 inH2O range
• Built-in pressure sensor, optional pneumotach flowmeter for measuring flow
• Pressure sensor and cylinder pressure sensor for correcting flow caused by compression
• Flow capacity of 16 liters/second.
• Capacity of 8.5 liters
• BTPS test can select heating and humidification
• Requires 115 or 230 VAC 50-60 Hz power supply
• Workbench tabletop size, 36 inches (91.5 cm) long, 14 inches (35.5 cm) wide, 7 inches (18 cm) high

(1) ATS simulation requires testing various waveforms for spirometers and peak flowmeters, and testing instrument accuracy
(2) Can set waveforms independently
(3) Real-time display of all test waveforms
(4) Includes pressure sensor, capable of simultaneously testing pressure changes during testing
(5) Option BTPS can simulate actual subject breathing conditions, with heating option to avoid water vapor condensation
(6) Includes 24 standard capacity waveforms and 26 flow rate waveforms tests for all ATS
(7) Test flow can reach 16L/S and above
(8) Test capacity can reach 8.5L and above
(9) Host size ≤ 91.5* 35.5* 18 cm
1120 Pulmonary Capacity Meter Check Waveform Generator Product Physical Picture: