The TIR 100-2 Infrared Emissivity Meter is a compact, handheld analytical instrument that uses non-destructive technology to measure thermal emissivity within seconds. This device is suitable for a wide range of materials, from low to high emissivity and from smooth to textured surfaces, even at elevated temperatures.
The measurement component, operational controls, control unit, and evaluation electronics are all space-savingly integrated into the handheld design. The instrument requires only a power source to operate, with user interaction facilitated through an illuminated touchscreen display.
Advantages of the TIR 100-2 Thermal Emissivity Meter:
- Measurements completed in seconds
- Easy operation and result viewing via intuitive touchscreen interface
- High measurement accuracy and repeatability
- Samples remain at ambient temperature during measurement
- Portable device suitable for use in the field or laboratory
Technical Specifications
- Model: TIR100-2
- Measurement range: 0.010–0.980
- Spectral range: 2.5 μm–50 μm
- Radiant source temperature: 100 °C
- Single measurement duration: approx. 5 s
- Measurement spot diameter: 5 mm
- Interface: USB-B
- Dimensions: 230 × 140 × 120 mm
- Main unit weight: 2 kg
- Total weight including packaging and calibration standards: 9 kg
- Power supply: 100–240 VAC, 50/60 Hz
- Operating environment: 15–30 °C, relative humidity <70%
- Accessories: aluminum protective case, calibration standards, power cable, user manual
The TIR100-2 infrared thermal emissivity meter comes standard with the following:
TIR100-2 model, with optional power voltage selection (115 V or 230 V).
Dual-in-one calibration standard: low-emissivity side of polished aluminum and high-emissivity side of blackbody cavity.
Calibration certificate for the standards, traceable to reference standards at Physikalisch-Technische Bundesanstalt (PTB), Berlin (GER), reference numbers 73213 PTB 16 and 73214 PTB 16. Validity period: two years.
Lockable portable hard-anodized aluminum case for easy storage and transportation of the TIR100-2 and calibration standards.
English-language printed manual.
Optional accessories for the TIR100-2 infrared thermal emissivity meter include: aluminum foil holder.
The foil clamp is a practical device that securely holds thin-coated foil samples (and other textiles) flat and evenly across the entire surface of the clamp plate, as shown in the figure.
Two magnetic clamps on each side secure the sample in place.
An additional benefit of using the foil holder is its ability to help maintain the sample at room temperature for a longer period during measurement. This is particularly useful when measuring thin materials such as coated foils, helping reduce thermal fluctuations and thereby improving the accuracy of emissivity measurements.