Rigel UNI-SiM handheld vital sign simulator from the UK is a 6-in-1 synchronous signal generator. It can simultaneously output six physiological signals - non-invasive blood pressure, blood oxygen, electrocardiogram, body temperature, invasive pressure, and respiration - in one go. It is battery-powered and portable, meeting the daily calibration and quality control requirements of hospital equipment departments and third-party metrology institutions for monitoring devices.
The Rigel UNI-SiM vital sign simulator is the world's first battery-powered six-in-one vital sign simulator, capable of conducting six simultaneous vital sign parameter tests.
The Rigel Uni-Sim life vital signs simulator can quickly, simply and accurately simulate NiBP, SpO2, ECG, temperature, IBP and respiratory functions. It is one of the most valuable tools available today.
In the SPO2 and NIBP simulations, new advanced technologies are utilized to reduce errors and minimize the testing time to the greatest extent. Manufacturer-specific simulation curves can be created and uploaded to further enhance accuracy.
The Bluetooth connection with various attachments and Med-eBase (a comprehensive database software) ensures that test results can be downloaded quickly and easily, the asset database can be managed, test sequences can be created, and test certificates can be generated.
Rigel UNI-SIM Vital Signs Simulator Complete Technical Specifications
Non-Invasive Blood Pressure (NIBP)
|
Parameter |
Specification |
| Simulation Waveform |
Oscillometric |
| Integrated Pump |
0 to 350 mmHg, user-configurable |
| Leak Test |
User-configurable between 0-350 mmHg |
| Timer |
Configurable up to 999 seconds |
| Digital Manometer |
0 - 410 mmHg |
| Pressure Accuracy |
+/- 0.5% Full Scale (FS) |
| Supported Pressure Units |
mmHg, inHg, kg/cm², cmH₂O, mBar, PSI, inH₂O, kPa |
Oxygen Saturation (SpO₂)
|
Parameter |
Specification |
| Measurement Range |
50 to 100% |
| Heart Rate Accuracy |
±1 bpm |
| Timer Function |
Test option for SpO₂ monitor response time verification |
Electrocardiogram and Arrhythmia Simulator (ECG)
|
Parameter |
Specification |
| ECG Configuration |
5-lead with advanced output capability |
| Basic Waveforms |
Normal Sinus Rhythm (NSR), ST Elevation, ST Depression, Myocardial Infarction, Tall T-wave |
| Heart Rate Range |
20 - 300 BPM, value input supported |
| Performance Waveforms |
Sine, Square, Triangle, Pulse |
| Pacemaker Waveforms |
Synchronized Atrial, Asynchronous Atrial |
| Standalone Pacemaker Modes |
Pacemaker pulse rate 60 BPM, ventricular pacing rate 70 BPM, atrial & ventricular pacemaker heart rate 70 BPM, R-wave detection |
| Atrial Arrhythmia Waveforms |
Sinus Arrhythmia (SA), Dropped Beats, Atrial Flutter (AFLT), Atrial Fibrillation (AFB), Paroxysmal Atrial Tachycardia (PAT), Junctional Premature Beats |
| Atrioventricular Conduction Abnormalities |
1st Degree AV Block, 2nd Degree AV Block - Mobitz Type I, 2nd Degree AV Block - Mobitz Type II, 3rd Degree AV Block, Right Bundle Branch Block (RBBB), Left Bundle Branch Block (LBBB), Left Anterior Fascicular Block |
| Ventricular Arrhythmia Waveforms |
Premature Ventricular Contractions (PVCs) - Intermittent, PVCs - Continuous, Bigeminy, Trigeminy, Ventricular Flutter (VFLT), Ventricular Fibrillation (VFB), Ventricular Tachycardia (VTC), Right Ventricular Outflow Tract PVCs |
Temperature Simulation (TEMP)
|
Parameter |
Specification |
| Simulation Mode |
YSI 400/700 static simulation |
| Preset Temperature Values |
25°C, 33°C, 37°C, 41°C |
Respiratory Simulation (RESP)
|
Parameter |
Specification |
| Respiratory Rate Options |
5, 10, 15, 30, 60, 120, 180 breaths per minute |
| Baseline Impedance Options |
250, 500, 750, 1000 ohms |
| Apnea Simulation |
0 - 60 seconds duration, 0 - 300 seconds interval |
Invasive Blood Pressure (IBP)
|
Parameter |
Specification |
| Number of Channels |
2 channels |
| Static Pressure Range |
0 to 300 mmHg, value input supported |
| Dynamic Pressure Range |
0 - 300 mmHg (systolic and diastolic pressure) |
General System Parameters
Operation Mode: Powered by a built-in rechargeable battery. Can operate independently without power supply.
Charger Specifications: 110-230VAC, 50/60Hz - universal voltage worldwide.
Battery Life: 4 to 24 hours (varies depending on the simulation mode)
Storage Capacity: Can store approximately 10,000 test records, facilitating batch detection and traceability
Communication Method: Bluetooth wireless data transmission, supporting computer online export of reports
Display: Monochrome 1/4VGA full graphic high-definition display, visual parameter operation
Operation Keyboard: Full alphanumeric keys, convenient and efficient parameter setting
Overall Weight: < 1.5kg
Body Dimensions: 270×110×75mm
Operating Environment: Temperature 10~30℃, humidity 0~90%RH (no condensation)
Storage Environment: -15℃ ~ +60℃
Protection Grade: IP40
Rigel UNI-SiM 6-in-1 Vital Signs Patient Simulator,NIBP SpO2 ECG Tester Product real photo